To verify the maximum efficacy of your CIDEXPLUS 28 Day Solution, ASP
recommends testing for the minimum effective concentration (MEC) prior to each
use. ASP offers test strips designed specifically for use with CIDEXPLUS
28 Day Solution. Easy to use and read, CIDEXPLUS 28 Day Solution Test Strips
are the most reliable way to verify that CIDEXPLUS 28 Day Solution will
destroy microorganisms including the highly resistant Mycobacterium tuberculosis.
For optimal results, please carefully follow Instructions for Use when using
test strips:
- Completely submerge indicated pad of the strip into solution. Hold for 1
second, then remove. Do not shake strip.
- Remove excess solution by standing the strip upright on a paper towel.
- Read results at 90 seconds. Do not read past 90 seconds. Pad will be completely
red to indicate effective solution. If any yellow remains on the pad apart
from the top line, the solution is ineffective and must be discarded.
- Tightly recap test strip bottle after each use.
- Discard any unused test strips 90 days after bottle is originally opened.
Download and print instructions for test strip use. Download
pdf